- Compact, wall-mounted transmitter for demand-controlled ventilation
- Incorporates the Vaisala CARBOCAP ® Sensor - the silicon-based NDIR sensor with unique internal referencing
- Advanced single-beam dual wavelength measurement with no moving parts
- Excellent long-term stability
|
Performance |
|
CO 2 measurement range Accuracy (including repeatability, non-linearity and calibration uncertainty)
|
0 … 2000 ppm
|
| Long-term stability |
± (2.5 % of range + 3 % of reading)
± 100 ppm CO 2 / 5 years |
| Response time T 90 |
1 min |
| Temperature dependence, typical |
-0.35 % of reading / °C |
| Pressure dependence, typical |
+0.15 % of reading/ hPa |
| Warm-up time |
1 min, 10 min for full specification |
| Product lifetime |
> 10 years |
Operating environment |
|
Temperature
|
-5 ... +45 °C (23 … 113 °F)
|
| Humidity |
0 ... 85 % RH |
| Pressure |
700 … 1200 hPa |
| Electromagnetic compatibility |
Complies with EMC standard EN61326-1:1997, +Am1:1998, Generic Environment |
Inputs and outputs |
|
Operating voltage
|
24 V (±20 %) AC/DC
|
| Power consumption |
<2 W |
| Outputs |
4 … 20 mA, 0 … 10 V, RS-485, 2-wire, non-isolated |
|
Product Brochure User Guide |